BIOSENSE WEBSTER INC LASSO® NAV ECO VARIABLE CATHETER; CATHETER, ELECTRODE RECORDING OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
D134301 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Cardiac Tamponade (2226)
|
| Date of Event |
03/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an (b)(6) male patient ((b)(6)) underwent an atrial fibrillation ablation and left atrial appendage occlusion (watchmen) procedure with a lasso® nav eco variable catheter and suffered cardiac tamponade requiring pericardiocentesis, surgical intervention and prolonged hospitalization.It was reported that during an atrial fibrillation and left atrial appendage occlusion watchmen case, a perforation to the left atrial appendage, and a pericardial effusion were noticed in the patient.There was a drop in blood pressure on the patient.The left atrial appendage perforation and the pericardial effusion were confirmed via the ice ultrasound system.The medical intervention provided to the patient was a pericardiocentesis.Some of the fluid removed was transfused back into the patient, but about 2000cc of fluid was removed.The procedure was continued, and they proceeded to place the watchmen device in the patient to close the leak.However, after being unsuccessful in closing the leak with the watchmen device, open heart surgery was performed on the patient.The patient was in stable condition at the time of reporting.The physician believed that the cause of the adverse event was procedure related due to the lasso catheter perforating the left atrial appendage while clocking it.The patient fully recovered with no residual effects.The patient required extended hospitalization because of the adverse event as the patient went to the icu due to cardiac surgery performed in response to the adverse event.A loaner generator was used and was already sent back; therefore, no information was available.The effusion was confirmed via the cardiothoracic surgeon, as a perforation in the left atrial appendage from the lasso catheter.The adverse event did not occur while ablating.A transseptal puncture was performed and prior to noting cardiac tamponade, ablation was performed.There was no evidence of a steam pop.The event occurred during the mapping phase.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned to biosense webster inc.(bwi) for evaluation.A visual inspection and an evaluation of all features of the device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.Per the event, several tests were performed.The magnetic and electrical features were tested, and no issues were observed.In addition, the product was deflecting and contracting correctly.No malfunctions were observed during the product analysis.A manufacturing record evaluation was performed for the finished device 30701559l number, and no internal actions related to the complaint were found during the review.No malfunction was observed during the product analysis.The root cause of the adverse event remains unknown.There may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.The ifu (instruction for use) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation or tamponade.The root cause of the adverse event remains unknown.There may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|