MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-82
Medical Device Problem Code Insufficient Cooling (1130)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
Livanova (b)(4) received a report that a heater-cooler system 3t had a defective cooling compressor.The issue was identified by the customer during routine maintenance.There was no patient involvement.
 
Additional Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).The customer reported that hole on the cooling coil was found.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Hospital has biomedical engineers.
 
Additional Manufacturer Narrative
H.10: through follow up communication livanova received confirmation that no patient was affected and it was learned that for this device regular maintenance was done according to the manufacturer ifu.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
See initial report.
 
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
H.10: the device was returned back to the manufacturer site for evaluation.The reported issue could be confirmed.In detail, it was found, that the cooling coil was leaking.Causing water to enter the cooling circuit.Due to this the fuse tripped.In addition, bearing damage was found on all pump motors.The complete cooling circuit and all pumps need to be replaced.The repair is not recommended from an economic perspective.It is likely, that the device will not be repaired.The reported event is a known issue.Livanova put in place corrective action to reduce this type of events in case of breach of the cooling coil being generated, due to stirrer flow in the tank leading to hole formation in a narrow area of the coil.However, in this specific case, the exact location of the hole on the cooling coil could not be determined.In october 2019, it was performed, the erosion kit upgrade of the device that includes the new design of the pump head of the stirrer motor to prevent the water flow directed to the narrow area of the cooling coil.The reported event was claimed after the upgrade.However, complaints database review revealed, a previous complaint from this customer on this specific unit.Reporting of nickel plating detachment.Thus, it is not possible to exclude that the condition of the cooling coil was already compromised, before the upgrade to such an extent, that the copper was rather degraded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Common Device Name
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14117005
Report Number9611109-2022-00172
Device Sequence Number11845793
Product Code DWC
UDI-Device Identifier04033817900511
UDI-Public(01)04033817900511(11)170302
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K191402
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number16-02-82
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer05/19/2022
07/27/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date06/16/2022
08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/02/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-