MAUDE Adverse Event Report: COVIDIEN NANOCROSS ELITE 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN NANOCROSS ELITE 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number AB14W060150150
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use the nanocross elite pta balloon during procedure to treat moderately calcified lesion in the proximal left superficial femoral artery(sfa).The vessel had little tortuosity and exhibited 80% stenosis.The artery diameter was 6mm and lesion length was 60mm.A 6fr sheath and non-medtronic guidewire were used.The balloon was inflated with a non-medtronic inflation device.There was no damage noted to packaging, i.E., shelf carton, hoop/tray and no issues noted when removing the device from the hoop/tray.No damage was noted to the balloon catheter.No kinks were noted in balloon catheter or wire.It was reported that both inflation and deflation issues were noted.Balloon partially inflated then wouldn¿t deflate.There were balloon inflation difficulties at 8atm.It was reported that there were balloon deflation difficulties.Device would not deflate at the lesion site.Deflation issues noted following first inflation.Deflation issues noted during subsequent inflation.The device did not pass through a previously deployed stent.No resistance was encountered when advancing the device.It took 15 minutes of backward pressure to get balloon deflated enough to go through the sheath.No patient injury.
 
Additional Manufacturer Narrative
Product analysis an indeflator with pressure gauge was used to inflate the balloon to nominal pressure 8 atms and rated burst pressure 14atms but the balloon would not inflate.Under a microscope, crimping marks were noted on the proximal balloon bond chamber medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOCROSS ELITE 0.14 OTW PTA DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14117274
Report Number2183870-2022-00142
Device Sequence Number4262064
Product Code LIT
UDI-Device Identifier00763000110505
UDI-Public00763000110505
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K132777
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2025
Device Model NumberAB14W060150150
Device Catalogue NumberAB14W060150150
Device Lot NumberB338824
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer10/13/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
Patient SexFemale
Patient Weight81 KG
-
-