• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. 4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. 4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number RFG-NT-2000
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the procedure, with the patient on the table, the generator screen was not responding and the case was cancelled.
 
Additional Manufacturer Narrative
One 4 lesion nt2000¿ pain management rf generator was received for analysis.Visual inspection revealed the input, output connectors and controls appeared to be free of physical damage.All the mounting hardware was secured.Normal wear was observed on the exterior enclosure.Ac power was applied to the device and powered on, it initialized and displayed the software application.The touchscreen was unresponsive and did not respond to tactile input.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, root cause of the field reported event was successfully isolated to abnormal functionality of the microprocessor within the upper assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR
Common Device Name
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14117444
Report Number2184149-2022-00101
Device Sequence Number2356337
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K111576
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-