• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx1761 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (refx1761) have been reported from the same facility.
 
Event or Problem Description
It was reported "stylet issue.It was severely kinked after picc was inserted.This particular patient was a failed catheter attempt, it would not advance.Pt was a 65 yo male with a right humerus fracture.Attempt was made to left arm, brachial vein, vein was a 0.5cm diameter at insertion site.".
 
Event or Problem Description
It was reported "stylet issue.It was severely kinked after picc was inserted.This particular patient was a failed catheter attempt, it would not advance.Pt was a 65 yo male with a right humerus fracture.Attempt was made to left arm, brachial vein, vein was a 0.5cm diameter at insertion site.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed; however, the root cause was not identified.The product returned for evaluation was two photographs which depicted a 5fr t/l provena powerpicc catheter.Both photographs depicted the distal end of the catheter, which appeared to have been trimmed just distal of the 42cm depth marking.Blood residue was visible on the sample.The 3cg tls stylet was visible protruding from the catheter tip.The stylet exhibited multiple kinks within the plastic-coated region.The stylet appeared to be intact.Stylet deformation was evident in the submitted photographs; however, inspection of the photographs was insufficient to identify the cause of the deformation.Consequently this complaint is confirmed as ¿cause unknown¿ at this time.Potential contributing factors include stylet manipulation and insertion of the stylet past the catheter tip during device insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14117553
Report Number3006260740-2022-01292
Device Sequence Number4250045
Product Code LJS
UDI-Device Identifier00801741155123
UDI-Public(01)00801741155123
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1395108D4
Device Lot NumberREFX1761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age65 YR
Patient SexMale
-
-