• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER 5F RADIOLOGY BASIC TRAY (DUAL-LUMEN); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER 5F RADIOLOGY BASIC TRAY (DUAL-LUMEN); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refr0313 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the powerpicc solo catheter was placed on (b)(6) 2022.The guidewire cannot be advanced smoothly and then it was found kinked in patient.The guidewire cannot be removed so the radiologist forcedly removed the guidewire from the peel away sheath introducer.Part of guidewire was broken and remained in patient¿s body.The remaining guidewire was removed under usg.Patient status/ intervention: the patients have serious svc stenosis (obstruction).
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a broken guidewire was confirmed.The product returned for evaluation was powerpicc solo.The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle.The returned product sample was evaluated and the guidewire was confirmed to be broken which allowed the outer coil wire to become unraveled.Microscopic examination of the fracture sites revealed the following: ¿ shearing of the wire at the break which was indicative of direct contact with a hard-edged instrument (such as an introducer needle); narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire; damage to the inside edge of the introducer needle which can occur if the guidewire is forcefully retracted against the needle, damaging the shape of the sharpened bevel; the weld tip of the wire was missing and could not be accounted for during the evaluation normal movement of the guidewire is away from the sharpened bevel and will not damage the wire.If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle.An examination of the wire structure revealed no potential damage/defect related to manufacture of the product.H3 other text: evaluation findings are in section h11.
 
Event or Problem Description
It was reported that the powerpicc solo catheter was placed on (b)(6) 2022.The guidewire cannot be advanced smoothly and then it was found kinked in patient.The guidewire cannot be removed so the radiologist forcedly removed the guidewire from the peel away sheath introducer.Part of guidewire was broken and remained in patient¿s body.The remaining guidewire was removed under usg.Patient status/ intervention: the patients have serious svc stenosis (obstruction).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO CATHETER 5F RADIOLOGY BASIC TRAY (DUAL-LUMEN)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14117635
Report Number3006260740-2022-01295
Device Sequence Number19892148
Product Code LJS
UDI-Device Identifier00801741028106
UDI-Public(01)00801741028106
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number3295135
Device Lot NumberREFR0313
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-