• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. ALT HA S CLR STD SZ 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. ALT HA S CLR STD SZ 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number ALT HA S CLR STD SZ 3
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: nv gxl liner neutral, 32mm id group 1 cups (cat# 130-32-51 / (b)(4)); biolox delta femoral head 32mm od, +0mm (cat# 170-32-00 / (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately four years post initial left tha, this (b)(6) female patient was brought back for left hip pain due to a loose stem.The hip was dislocated, the femoral head was removed, attention turned to the cup, the liner was removed with the help of an osteotome, the cup was checked and found to be well fixed.A new xle g1 lipped liner was implanted.The stem was checked, found to be loose and removed.A new stem and head was then implanted.Patient was revised to xle g1 32mm lipped insert, a std 7 taper wedge stem and a 32++0 biolox head.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.The device will be return.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the femoral stem.The most probable root cause associated with the reported event of "loosening - femoral stem¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALT HA S CLR STD SZ 3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14117992
Report Number1038671-2022-00416
Device Sequence Number2291657
Product Code JDI
UDI-Device Identifier10885862321053
UDI-Public10885862321053
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K162732
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2022
Device Model NumberALT HA S CLR STD SZ 3
Device Catalogue Number190-30-03
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/23/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-