• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number H700488
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
Related manufacturing ref: 2184149-2022-00104, 2184149-2022-00105.During an atrioventricular nodal reentrant case, the user switched from cryo to rf and received a 999-impedance error on the generator.Patch placement was confirmed and replaced with good communication between the generator and amplifier.The cables were inspected, reseated, rebooted the amplifier, generator, tactisys and the dws which did not resolve the issue.The catheter was replaced with one from a different lot and attempted a second se port, but the issue remained.The case continued by switching back to cryo with no patient harm.
 
Additional Manufacturer Narrative
One ampere¿ rf ablation generator was received for evaluation.Evaluation of the logs revealed an ¿lag processing 1 notifications for cycle¿ which confirms the reported symptom.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.Based on the investigation and information provided to abbott, this symptom was not able to be duplicated, however, the root cause is attributed to software design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPERE¿ RF ABLATION GENERATOR
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14118078
Report Number2184149-2022-00103
Device Sequence Number2276962
Product Code LPB
UDI-Device Identifier05415067004888
UDI-Public05415067004888
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P110016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH700488
Device Catalogue NumberH700488
Device Lot Number4798977
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2022
Initial Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENSITE X EP SYSTEM AMPLIFIER; TACTISYS¿ QUARTZ ABLATION SYSTEM
Patient SexUnknown
-
-