MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 3549211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Post Operative Wound Infection (2446)
Date of Event 02/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: infection.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent a breast reconstruction primary with mentor cpx 4 breast tissue expander with suture tabs 275cc and experienced infection on the left side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent removal on an unspecified date.
 
Additional Manufacturer Narrative
On apr 16 2022, additional information was received indicating the date of explantation was (b)(6) 2022.The device serial number was identified as (b)(6).On apr 26 2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A photo of the device was returned for evaluation.Photo evaluation summary: upon visual evaluation of the image provided in the complaint, no apparent damage or visual anomalies were observed.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the tissue expander.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the tissue expander is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned tissue expander.During the visual evaluation, no apparent damage or visual anomalies were observed on the med height te w/sut tabs 275cc tissue expander.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
6465917981
MDR Report Key14118093
Report Number1645337-2022-04253
Device Sequence Number2277519
Product Code LCJ
UDI-Device Identifier00081317006718
UDI-Public00081317006718
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/18/2025
Device Model Number3549211
Device Catalogue Number3549211
Device Lot Number9522318
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer04/16/2022
05/25/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date05/05/2022
06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-