• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOBIUS MOBILITY LLC. IBOT PMD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOBIUS MOBILITY LLC. IBOT PMD Back to Search Results
Model Number IBOT PMD
Health Effect - Clinical Codes Laceration(s) (1946); Concussion (2192)
Date of Event 03/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Mobius mobility investigation: the device involved in the incident was not returned to mobius mobility, however the event logs, fault code data, and black box data were retrieved by the mobius mobility service team.The homeowner's security camera footage was also provided to mobius mobility for review.A failure investigation was performed using the data and video footage.The data from the device shows that 55 seconds after entering stair mode, an event occurred in which the device went into cluster safety lock and exceeded its pitch limit.Cluster safety lock is the device's detection of a loss of pitch control by the user or assistant during stair climbing.Prior to this event, the data shows a portion of step descent with adequate pitch control (braking), then a 5 second period of forward pitch input that causes rapid device movement in the descending direction.The pitch is not adequately controlled by the assistant, resulting in the event.The device data logs are consistent with the reported incident and the user's video.No other alarms were present in the logs that would have contributed to the event.The cause of the incident is most attributed to the assistant's improper foot placement (stance) on the 1st step down from the landing, as shown in the video, which did not allow for proper pitch control of the device.Following the incident, the logs show the device was powered off/on and continued to operate normally with "service required" flag set, which limits device functionality disabling stair mode.This is expected device behavior after the device exceeds its pitch limit.Thus, the device did not malfunction and performed as expected.
 
Event or Problem Description
On 03/29/2022 the user called mobius mobility to report an incident involving the ibot® pmd which occurred on (b)(6) 2022.The user was seated in the ibot® pmd while being assisted down the steps by his wife, who has passed mobius mobility's assisted stair training, and had an event resulting in him being ejected from the device.User stated he was not wearing the provided lap belt at the time of the event as advised in accompanying device labeling.The user reported he sustained a concussion and a laceration on his tongue which required stitches.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBOT PMD
Common Device Name
IBOT
Manufacturer (Section D)
MOBIUS MOBILITY LLC.
540 north commercial st.
suite 310
manchester NH 03101
Manufacturer (Section G)
MOBIUS MOBILITY LLC.
540 n. commercial street
suite 310
manchester NH 03101
Manufacturer Contact
joseph sullivan
540 n. commercial street
suite 310
manchester, NH 03101
6034258703
MDR Report Key14118348
Report Number3014522447-2022-00001
Device Sequence Number2428195
Product Code IMK
UDI-Device Identifier00857584008003
UDI-Public00857584008003
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K172601
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberIBOT PMD
Device Catalogue NumberMMPX-10000-000
Device Lot NumberN/A
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Initial Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexMale
Patient Weight118 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-