| Model Number |
8888256529 |
| Medical Device Problem Codes |
Complete Blockage (1094); Decrease in Suction (1146)
|
| Health Effect - Clinical Code |
Abdominal Distention (2601)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reports: there is an issue with the replogle suction catheter.The nursing staff tried one this morning and it was defective.Additional information was received on 4/14/22 stating that the replogle was found occluded; it would not aspirate or flush with air.The device was replaced and no further medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed, and no discrepancy was found according to the reported condition.There were no samples or photographs submitted for review and investigation.Based on the limited information received, the reported complaint could not be confirmed nor related to a manufacturing process.All associated lot documentation was carefully reviewed with no issues or discrepancies were found related to the reported complaint.Based on the analysis, the observed occurrence of harm improbable which is below the range of the expected occurrence rate therefore corrective actions are not deemed necessary at this time.
|
| |
|