Model Number BF-UC180F
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device was sent to an olympus service center for evaluation.Inspection and testing confirmed the reported event.The biopsy port was loose due to physical stress.In addition, bending angle did not meet specification and the adhesive on the bending section cover had deterioration and separation of the thread-wound sections.The investigation is ongoing.Therefore; the root cause of the reported event cannot be determined at this time.However; if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported the metal part at the biopsy port on the scope had come off.Additional details were requested regarding the reported event.No additional information was provided.There was no harm or user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been nearly 5 years since the subject device was manufactured.Based on the results of the investigation, the specific root cause of the damaged biopsy port could not be determined at this time.The following information is stated in the instructions for use which may have prevent the event: "do not apply excessive force to or bend the distal end of the endoscope.Instrument damage may occur." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Common Device Name
ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14118642
Report Number8010047-2022-06428
Device Sequence Number2276978
Product Code PSV
UDI-Device Identifier04953170341823
UDI-Public04953170341823
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K070983
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-UC180F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-