MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER Back to Search Results
Model Number 998806
Medical Device Problem Codes Material Puncture/Hole (1504); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that when the user tried to inflate the balloon of balloon dilation catheter at the stenosis, it did not inflate.When the person in charge of the agency visually checked after collecting the actual product, it was found that there was a hole in the shaft part in front of the balloon.Although, it was stated that there was a hole in the shaft part, doctors and nurses who were at the site at the moment had commented on "possibility of operation in the procedure when inserting the guide wire".
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.Photo samples show a tear in the catheter shaft.The balloon dilation catheter was returned with the original packaging.The device was visually inspected at 12"-18" under normal room lighting and a cut was noted on the outer shaft.A white crystal-like substance was noted on and inside the catheter, possibly due to decontamination process or contrast fluid.An in-house 0.038" guidewire as inserted and passed freely per specifications.The device was inspected under 7-10x magnification to better visualize the hole; it was noted to be approximately 0.50" from the proximal end of the balloon assembly.The hole appears to be a cut in the outer shaft made by a sharp object.A portion of the outer shaft was dissected away from the whole device and the inner shaft was removed.No defects were noted to the inner shaft.A potential root cause for this event could be, "shaft wall too thin".The device was used for treatment purposes.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following."the x-force® balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident.Caution: if resistance is felt when removing either the catheter or the guidewire from the introducer sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation.Store the balloon dilation catheter in a cool, dry, dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials." corrections: f,h.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that when the user tried to inflate the balloon of balloon dilation catheter at the stenosis, it did not inflate.When the person in charge of the agency visually checked after collecting the actual product, it was found that there was a hole in the shaft part in front of the balloon.Although, it was stated that there was a hole in the shaft part, doctors and nurses who were at the site at the moment had commented on "possibility of operation in the procedure when inserting the guide wire".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14119110
Report Number1018233-2022-02543
Device Sequence Number14026072
Product Code EZN
UDI-Device Identifier10801741052771
UDI-Public(01)10801741052771
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/30/2023
Device Model Number998806
Device Catalogue Number998806
Device Lot NumberBMFTFM55
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-