Model Number TT012
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo images were received and are pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.No lot or batch number was provided therefore a device history could not be done.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a thoracoscopic pneumonectomy, the sleeve was broken off when the trocar was removed from the patient after the surgery.The broken pieces were retrieved.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 6/2/2022.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the sleeve tt012 device was received with the tip of the sleeve broken.The event reported was confirmed and it is related to improper use of the device.One possible cause for the damage found on the sleeve may be excessive external load placed on the device.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 5/11/2022.Photo analysis: this is an analysis for a photo submitted to ethicon endo surgery for evaluation.During the visual analysis, the following was observed: the photo shows the endopath*thoracic trocar sleeve broken on the tip.Based on the photo review, the event describe is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ENDOPATH*THORACIC TROCAR SLEEV
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
MDR Report Key14119124
Report Number3005075853-2022-02424
Device Sequence Number14026227
Product Code GEA
UDI-Device Identifier10705036013372
UDI-Public10705036013372
Combination Product (Y/N)N
PMA/510(K) Number
K920110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTT012
Device Catalogue NumberTT012
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer04/19/2022
05/13/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date05/11/2022
06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-