| Model Number |
TT012 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo images were received and are pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.No lot or batch number was provided therefore a device history could not be done.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a thoracoscopic pneumonectomy, the sleeve was broken off when the trocar was removed from the patient after the surgery.The broken pieces were retrieved.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/2/2022.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the sleeve tt012 device was received with the tip of the sleeve broken.The event reported was confirmed and it is related to improper use of the device.One possible cause for the damage found on the sleeve may be excessive external load placed on the device.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/11/2022.Photo analysis: this is an analysis for a photo submitted to ethicon endo surgery for evaluation.During the visual analysis, the following was observed: the photo shows the endopath*thoracic trocar sleeve broken on the tip.Based on the photo review, the event describe is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.
|
| |
|