ROCHE DIAGNOSTICS COAGUCHEK VANTUS; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number COAGUCHEK INRANGE
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Handling Problem (3265)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation is patient/consumer.The meter and strips were requested for investigation.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of a software issue with a coaguchek vantus meter.The meter showed a duplicate result in the meter memory.The patient only tested one time on (b)(6) 2022.The meter memory showed the following results.On (b)(6) 2022 the meter result was 4.3 inr at 15:57.On (b)(6) 2022 the meter result was 4.3 inr at 16:03.The result was reported and the patient was advised to hold warfarin for 2 days.There were no negative effects from holding medication.The patient's therapeutic range is 2.5-3.5 inr.The test strip lot number is 49682521 with an expiration date of 30-apr-2022.
 
Additional Manufacturer Narrative
The reporter's meter was provided for investigation where they were tested using retention strips and controls.Testing results (qc range = 4.1 ¿ 6.8 inr): qc 1: 5.3 inr, qc 2: 5.3 inr, qc 3: 5.2 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The meter did not run slow.The results measured on (b)(6) 2022 were measured on the meter 46 minutes apart and there were two different entries as expected in the strip counter (nb.86 and 87).Medwatch fields d9 and h3 have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK VANTUS
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key14119330
Report Number1823260-2022-01065
Device Sequence Number4334559
Product Code GJS
UDI-Device Identifier00365702700499
UDI-Public00365702700499
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170960
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCOAGUCHEK INRANGE
Device Catalogue Number07729952160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-