• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD PRESET¿ SYRINGE WITH ATTACHED NEEDLE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD PRESET¿ SYRINGE WITH ATTACHED NEEDLE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 364314
Device Problems Melted (1385); Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd had not received samples, but 2 photos were provided for investigation.The photos were reviewed and the indicated failure mode for deformed plunger stopper was observed.The failure mode of missing needle shield was not observed from the photos.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode deformed plunger stopper.This complaint has been unconfirmed for the failure mode missing needle shield.Bd determined that the root cause of the deformed plunger stopper was attributed to worn manufacturing equipment as well as interfering parts within the manufacturing process.Several corrective actions are being put in place which include repositioning equipment and replacing worn parts.The process will continue to be reviewed after corrective actions have been completed.Bd was not able to identify a root cause for the missing needle shield.
 
Event Description
It was reported when unpacking the bd preset¿ syringes with attached needles there was deformed plungers and sterility issues.This event occurred 3 times.The following information was provided by the initial reporter.The customer stated: "when unpacking, two of the plunger stoppers were melted and one needle shield was missing.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRESET¿ SYRINGE WITH ATTACHED NEEDLE
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14119345
MDR Text Key289456155
Report Number9617032-2022-00365
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K022426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2023
Device Catalogue Number364314
Device Lot Number1197895
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-