• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA APDL; TUBE, DRAINAGE, SUPRAPUBIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA APDL; TUBE, DRAINAGE, SUPRAPUBIC Back to Search Results
Model Number 30935
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 05apr2022.A flexima all purpose drainage catheter was returned for analysis with no reported complaint.However, device analysis revealed the stent had detached.
 
Additional Manufacturer Narrative
Device eval by mfr: the catheter was returned with the metal cannula and trocar bent and loaded.A tear and detachment were observed in the pigtail section of the stent.However, the detached pigtail section was still held together by the suture string.The cannula was unloaded without issues.The metal cannula was unloaded from the catheter and no resistance was felt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA APDL
Common Device Name
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14120354
Report Number2134265-2022-04435
Device Sequence Number2277580
Product Code FFA
UDI-Device Identifier08714729323099
UDI-Public08714729323099
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K944290
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2024
Device Model Number30935
Device Catalogue Number30935
Device Lot Number0028514203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2022
Initial Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-