| Model Number |
30935 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 05apr2022.A flexima all purpose drainage catheter was returned for analysis with no reported complaint.However, device analysis revealed the stent had detached.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by mfr: the catheter was returned with the metal cannula and trocar bent and loaded.A tear and detachment were observed in the pigtail section of the stent.However, the detached pigtail section was still held together by the suture string.The cannula was unloaded without issues.The metal cannula was unloaded from the catheter and no resistance was felt.
|
| |
|