BOSTON SCIENTIFIC CORPORATION FLEXIVA (TM) TRACTIP HIGH POWER SINGLE-USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068403961
Medical Device Problem Codes Overheating of Device (1437); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a flexiva tractip high power single use laser fiber was used during a procedure, performed on (b)(6) 2022.During the procedure, the laser fiber was not working and the connector was warm to the touch.The procedure was completed with another flexiva tractip high power single use laser fiber.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.
 
Additional Manufacturer Narrative
Block h6: medical device problem code a1005 captures the reportable event of fiber connector overheats.Block h10: investigation results the returned flexiva tractip high power single use laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in one piece.The fiber body measured 314.3 cm from the distal tip to the proximal end of the sma connector.The exposed glass tip measured 4 mm and had normal degradation.A functional evaluation revealed that the light from the sma connector was distorted, indicating that there was fracture within the connector.No other problems with the device were noted.The reported event of the flexiva tractip high power single use laser fiber connector overheats was confirmed.It is likely that procedural handling of the device during set-up or use contributed to the fracture within the fiber connector.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Additionally, the ifu states, in the event of reduced or no laser energy output during application the fiber connector may be hot to touch.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6), 2022 that a flexiva tractip high power single use laser fiber was used during a procedure, performed on (b)(6), 2022.During the procedure, the laser fiber was not working and the connector was warm to the touch.The procedure was completed with another flexiva tractip high power single use laser fiber.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA (TM) TRACTIP HIGH POWER SINGLE-USE LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
COHERENT INCORPORATED
5100 patrick henry drive
santa clara CA 95054
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14121440
Report Number3005099803-2022-02079
Device Sequence Number2356465
Product Code GEX
UDI-Device Identifier08714729808176
UDI-Public08714729808176
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K110685
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2024
Device Model NumberM0068403961
Device Catalogue Number840-396
Device Lot Number0000011437
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-