SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD SMARTSITE¿ EXTENSION SETS; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 20041E
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 10 bd smartsite¿ extension sets were difficult to disconnect from the device.The following information was provided by the initial reporter: "can not be separated from the screw connector of the intravenous infusion set.".
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6: investigation summary: a 20041e product was not available for investigation; however the customer confirmed that the complaint sample was from lot: 21065755.As part of the investigation, the customer provided a photograph of the affected sample; analysis of the photograph indicated that the female luer of the 20041e product could not be disconnected from the male luer of a pratt & whitney infusion set.The details of this feedback were forwarded to the manufacturing site for investigation.A review of the production records for lot: 21065755 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.The root cause of the customer¿s experience could not be determined as the sample was not available for investigation.In this instance, without a sample, it is not possible to determine whether a manufacturing defect could have caused or contributed to the customer¿s experience.A review of the customer feedback database indicates that this is an isolated feedback with no further reports of this nature against the 20041e product over the past 12 months.
 
Event or Problem Description
It was reported that 10 bd smartsite¿ extension sets were difficult to disconnect from the device.The following information was provided by the initial reporter: "can not be separated from the screw connector of the intravenous infusion set.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SMARTSITE¿ EXTENSION SETS
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14121802
Report Number9616066-2022-00390
Device Sequence Number16713613
Product Code FPA
UDI-Device Identifier27613203021665
UDI-Public27613203021665
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/15/2024
Device Model Number20041E
Device Catalogue Number20041E
Device Lot Number21065755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date04/15/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-