| Model Number |
20041E |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that 10 bd smartsite¿ extension sets were difficult to disconnect from the device.The following information was provided by the initial reporter: "can not be separated from the screw connector of the intravenous infusion set.".
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: a 20041e product was not available for investigation; however the customer confirmed that the complaint sample was from lot: 21065755.As part of the investigation, the customer provided a photograph of the affected sample; analysis of the photograph indicated that the female luer of the 20041e product could not be disconnected from the male luer of a pratt & whitney infusion set.The details of this feedback were forwarded to the manufacturing site for investigation.A review of the production records for lot: 21065755 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.The root cause of the customer¿s experience could not be determined as the sample was not available for investigation.In this instance, without a sample, it is not possible to determine whether a manufacturing defect could have caused or contributed to the customer¿s experience.A review of the customer feedback database indicates that this is an isolated feedback with no further reports of this nature against the 20041e product over the past 12 months.
|
| |
|
Event or Problem Description
|
|
It was reported that 10 bd smartsite¿ extension sets were difficult to disconnect from the device.The following information was provided by the initial reporter: "can not be separated from the screw connector of the intravenous infusion set.".
|
| |
|