| Model Number |
Z9002 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age or dob, weight, & ethnicity unknown, not provided.If implanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.If explanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) was fully inserted and had to be removed/replaced with a back up lens because the iol was torn.The replacement lens is of a same model and diopter.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: upon further review it was noted that additional info.Received such as dr.(b)(6), patient dob (date of birth) (b)(6) 1947, os (left eye) affected.Field below updated.Section a2: date of birth: (b)(6), 1947.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9.Device available for evaluation? yes.Returned to manufacturer on: apr 14 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification was performed and revealed that the lens was cut into 3 pieces and both haptics were damaged.Due to the returned condition of the lens (cut into 3pieces) no further product evaluation can be performed.The complaint issue dc-iol torn was not confirmed.The other observed issue during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no additional complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|