|
SMITH & NEPHEW, INC. SMART TSF CONVERSION CLAMP; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION
|
Back to Search Results |
|
| Catalog Number |
71078550 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during surgery, a smart tsf conversion clamps broke outside the patient as it was being tightened on tsf ring.Surgery was resumed, after a non-significant delay, with a change in the surgical technique.Patient was not injured as consequence of this problem ((b)(4)) three additional clamps were received, but it is unspecified what their failures are ((b)(4)).
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The head of the bolt is broken, causing the internals of the device to eject.This renders the device inoperative.The device shows signs of extensive use.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: case-2021-00080145-2.
|
| |
|
Search Alerts/Recalls
|
|
|