MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR PLUS C-FLEX CPAP SYSTEM; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR PLUS C-FLEX CPAP SYSTEM; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 261P
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
A ventilator was returned for service.There was no harm or injury reported.During the evaluation at the manufacturer's service center, the sound abatement foam was replaced for repair.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported death, life-threatening, which was marked incorrectly as this is product problem only as mentioned in previous report.Incorrectly marked 30 days and initial including 15 days, which should marked as 30 days and initial only in this follow up report.Incorrectly marked in previous report, which is correctly marked as consumer and foreign in this follow up report.
 
Additional Manufacturer Narrative
The device was retuned to the manufacturer and tobacco contamination observed during device evaluation.Device available for evaluation, initial reporter name and address, pma/510k, device evaluated by mfr were updated/corrected in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR PLUS C-FLEX CPAP SYSTEM
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key14132232
Report Number2518422-2022-11354
Device Sequence Number16717137
Product Code BZD
UDI-Device Identifier00606959028654
UDI-Public00606959028654
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number261P
Device Catalogue Number261P
Was Device Available for Evaluation? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 10/07/2015
Supplement Date Received by Manufacturer10/07/2015
08/23/2022
Initial Report FDA Received Date04/17/2022
Supplement Report FDA Received Date04/18/2022
08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-