• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Unintended Application Program Shut Down (4032)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2022
Event Type  Injury  
Event Description
It was reported that, during machine set up for a cori assisted surgery, on clicking "begin operation" the error "warning internal error: the system has detected that the application unexpectedly exited" popped on the screen.The machine was restarted and the problem still persisted.The built software was reloaded and the problem still persisted.The entire data was wiped out using sd0091 and then the build software and idb was installed and the problem still persisted.On checking the log files it was found that there was some communication error between cori and tcu (tool controller unit) and it is resulted in internal error.The procedure was completed by switching to navio without significant delays.No patient injuries were reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H3, h6: the real intelligence cori, pn: rob10024, sn: (b)(6) used for treatment was not returned to the designated complaint unit for evaluation, therefore visual and functional inspections could not be performed.The software files were downloaded from the device and provided for investigation.The xsession log was reviewed and found that there is a communication hardware (ethernet) issue that is preventing the cori host computer and tcu from communication.Contributing factors for this ethernet issue could be a connectivity issue within the ethernet cable between the computer and the tcu, or the ethernet port(s) of the tcu and/or host computer.In the event of a system failure, refer to the recovery procedure guidelines in the real intelligence cori for knee arthroplasty user manual.A failure can consist of, but is not limited to, a system software crash, unrecoverable hardware failure, handpiece failure with no backup available, tracker failure or loss of contact with bone that is unrecoverable, etc.This situation is captured in the risk assessment released at the time of the complaint.The failure mode and associated risk have been anticipated within the risk file.As a result of the risk assessment the documented risk file will undergo further investigation by the site quality team.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review found similar reports.A review of prior escalation actions was performed and found no actions that are applicable to the scope of the reported complaint.This issue will be continuously monitored through complaint investigation and post market surveillance.Based on the investigation, no containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14132541
MDR Text Key289430608
Report Number3010266064-2022-00299
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
-
-