It was reported that, during machine set up for a cori assisted surgery, on clicking "begin operation" the error "warning internal error: the system has detected that the application unexpectedly exited" popped on the screen.The machine was restarted and the problem still persisted.The built software was reloaded and the problem still persisted.The entire data was wiped out using sd0091 and then the build software and idb was installed and the problem still persisted.On checking the log files it was found that there was some communication error between cori and tcu (tool controller unit) and it is resulted in internal error.The procedure was completed by switching to navio without significant delays.No patient injuries were reported.
|
H3, h6: the real intelligence cori, pn: rob10024, sn: (b)(6) used for treatment was not returned to the designated complaint unit for evaluation, therefore visual and functional inspections could not be performed.The software files were downloaded from the device and provided for investigation.The xsession log was reviewed and found that there is a communication hardware (ethernet) issue that is preventing the cori host computer and tcu from communication.Contributing factors for this ethernet issue could be a connectivity issue within the ethernet cable between the computer and the tcu, or the ethernet port(s) of the tcu and/or host computer.In the event of a system failure, refer to the recovery procedure guidelines in the real intelligence cori for knee arthroplasty user manual.A failure can consist of, but is not limited to, a system software crash, unrecoverable hardware failure, handpiece failure with no backup available, tracker failure or loss of contact with bone that is unrecoverable, etc.This situation is captured in the risk assessment released at the time of the complaint.The failure mode and associated risk have been anticipated within the risk file.As a result of the risk assessment the documented risk file will undergo further investigation by the site quality team.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review found similar reports.A review of prior escalation actions was performed and found no actions that are applicable to the scope of the reported complaint.This issue will be continuously monitored through complaint investigation and post market surveillance.Based on the investigation, no containment or corrective actions are recommended at this time.
|