• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE FOOT PEDAL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE FOOT PEDAL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10026
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during the femur mapping stage in a cori assisted surgery, the error "foot pedal cable disconnected" popped on screen.After resetting the foot pedal connection, the error "pedal cable disconnected" appeared during the femur cut stage.The surgery was completed with manual instrumentation with a significant delay.The patient was not harmed beyond the reported problem.
 
Manufacturer Narrative
H3, h6: the real intelligence foot pedal, part number rob10026, serial number (b)(6), used for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.The software files were downloaded from the device and provided for investigation.When the foot pedal is connected a voltage spike is generated, the log files show a voltage above the normal which results in a error.The symptoms relating to the problem can be compared to bug 1973.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The real intelligence cori for knee arthroplasty user manual (500230 rev d) provides guidelines for recovering to a fully manual procedure in the "recovery procedure guidelines, appendix d".The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.A historical review concluded that no prior escalation actions are applicable to the scope of the reported complaint.The most likely cause of this event is associated with a software bug.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE FOOT PEDAL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14132549
MDR Text Key289430233
Report Number3010266064-2022-00300
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757444
UDI-Public00885556757444
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10026
Device Catalogue NumberROB10026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2022
Initial Date FDA Received04/18/2022
Supplement Dates Manufacturer Received02/17/2023
Supplement Dates FDA Received02/20/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ROB20000/CORI ROBOTICS USA/SN (B)(6)
Patient Outcome(s) Hospitalization; Other;
-
-