This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c-us is available in the usa with a 510k number k190805.We checked the returned unit and confirmed that the image guide fiber bundle (cfb) fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the image guide fiber bundle (cfb).In addition, we confirmed that the angle wire fluid damage, the segment fluid damage, the u/d and the r/l pulley wires fluid damage, the light guide fiber bundle (lcb) fluid damage, the control body corroded, and the insertion flexible tube (ift) leak; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|