MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant products: product id: 8709, serial#: (b)(4), implanted: (b)(6) 2009, explanted: (b)(6) 2022, product type: catheter.Other relevant device(s) are: product id: 8709, serial/lot #: (b)(4), ubd: 22-apr-2011, udi#: (b)(4).Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving gablofen (1000 mcg/ml at 399 mcg/day) via an implanted pump.It was reported that the patient was in for a normal eri (elective replacement indicator) pump change.The physician noted that the dose had slowly increased over time, but the patient had no signs of withdrawal.Cap (catheter access port) aspiration was attempted on (b)(6) 2022, but they were not able to aspirate.Intraoperatively the catheter dripped spontaneously but aspiration was difficult, so the hcp decided to replace the pump and the catheter.There were no environmental, external, or patient factors that may have led or contributed to the issue.The issue was noted to be resolved, and it was indicated that the hcp had no further information to provide regarding the event.
|
|
Manufacturer Narrative
|
H3: analysis of catheter (s/n: (b)(6)) identified no anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|