• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. REAMER, LOW PROFILE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. REAMER, LOW PROFILE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number REAMER, LOW PROFILE
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022, it was reported by a sales representative via sems that the tip of an ar-1408lp reamer broke inside the patient, the patient was not affected or harmed.The surgeon was able to remove the pieces from the patient and the case was completed successfully.This was discovered during use with no impact to the patient.Additional information has been requested.On (b)(6) 2022, the sales representative provided the following information via email: this event occurred during a distal biceps repair procedure on (b)(6) 2022.The case was completed as stated and the surgeon was happy with the repair.The bone quality of the patient was good.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation identified that the tips of the reamer had fractured off during the procedure.Device testing could not be performed due to the damaged of the device.The cause of the reported failure mode is undetermined.However, the most likely probable cause is applying excessive mechanical forces upon insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER, LOW PROFILE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14135436
Report Number1220246-2022-04736
Device Sequence Number4706098
Product Code LXH
UDI-Device Identifier00888867016729
UDI-Public00888867016729
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberREAMER, LOW PROFILE
Device Catalogue NumberAR-1408LP
Device Lot Number10908319
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-