U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS BIOFLO PICC WITH ENDEXO TECHNOLOGY; BIOSTABLE PICC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS BIOFLO PICC WITH ENDEXO TECHNOLOGY; BIOSTABLE PICC Back to Search Results
Model Number 45876
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
A senior clinical risk advisor reported an issue with a bio-stable 4f sl 55cm mst-70 kit.Some time after placement of the device, the picc line tubing developed a hole in the sterile, dressed portion of the tubing; thus rendering the line ineffective.It is unknown how or when the hole occurred; however, the picc was removed and a peripheral intravenous line (piv) had to be inserted.This inhibited the patient from receiving tpn.The patient did not experience any adverse effects or as a result of this incident.
 
Additional Manufacturer Narrative
It was reported that the device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint of hole in extension leg was not confirmed as no product was returned for evaluation.Without receiving product for evaluation a root cause for this event cannot be determined.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: as noted during the review of the directions for use (dfu) that is supplied in the reported kit, the following precautions/warnings are stated: warnings: do not use if package is opened or damaged.Do not fully insert catheter up to suture wing.Do not use sharp objects to open package as damage to the device may occur.If catheter and accessories show any sign of damage (crimped, crushed, cut, etc.), do not use.Do not use sharp instruments near the extension tubes or catheter shaft.Precaution: do not use sharp objects to open package.Catheter repair: in the event that the catheter is accidentally torn or broken, it is recommended that the catheter be replaced.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO PICC WITH ENDEXO TECHNOLOGY
Common Device Name
BIOSTABLE PICC
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key14135570
Report Number1317056-2022-00071
Device Sequence Number17700044
Product Code LJS
UDI-Device IdentifierH965458760
UDI-PublicH965458760
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number45876
Device Catalogue Number45876
Device Lot Number5663889
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-