ARTHREX, INC. REAMER, LOW PROFILE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number REAMER, LOW PROFILE
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2022, it was reported by a sales representative via sems that an ar-1409lp reamer, low profile broke when reaming the tibia.Hard bone caused the wings of the reamer to break off when reamer impacted tibial bone when reaming over guide wire.Wings could not be located.Post op x-rays showed metal pieces in patients leg medial to tibial tunnel.On 04/06/2022, the sales representative stated the following information via email: this event occurred during an acl reconstruction procedure on (b)(6) 2022.Another reamer was used to complete the case.The patient is fine to date and has no effect recorded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER, LOW PROFILE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14135718
Report Number1220246-2022-04744
Device Sequence Number4706100
Product Code LXH
UDI-Device Identifier00888867016767
UDI-Public00888867016767
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberREAMER, LOW PROFILE
Device Catalogue NumberAR-1409LP
Device Lot Number14331267
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Initial Report FDA Received Date04/18/2022
Date Device Manufactured10/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-