• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. CHECKPOINT, 3.5 HEX IMPACTION STERILE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. CHECKPOINT, 3.5 HEX IMPACTION STERILE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 111653
Medical Device Problem Codes Use of Device Problem (1670); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
As reported: "checkpoint was left in greater troch and removed post op.Surgical delay > 30 minutes.Case type / application: tha" spoke to mps.After surgery was complete and the patient was out of the o.R., post-op x-ray showed the checkpoint left in the patient.Patient was brought back into the o.R., put back under anesthesia, and the checkpoint was removed.There was no intra-operative delay, mps was trying to report that the removal took > 30 minutes.Mps may be able to provide some additional patient information.
 
Event or Problem Description
As reported: "checkpoint was left in greater troch and removed post op.Surgical delay > 30 minutes.Case type / application: tha" spoke to mps.After surgery was complete and the patient was out of the o.R., post-op x-ray showed the checkpoint left in the patient.Patient was brought back into the o.R., put back under anesthesia, and the checkpoint was removed.There was no intra-operative delay, mps was trying to report that the removal took > 30 minutes.Mps may be able to provide some additional patient information.
 
Additional Manufacturer Narrative
Reported event: an event regarding other involving a mako checkpoint was reported.The event was not confirmed because the product was not available for inspection.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there are no other complaints with a similar failure mode for this lot.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.H3 other text : not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHECKPOINT, 3.5 HEX IMPACTION STERILE
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14135862
Report Number3005985723-2022-00047
Device Sequence Number4142336
Product Code OLO
UDI-Device Identifier00848486017316
UDI-Public00848486017316
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K112507
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number111653
Device Catalogue Number111653
Device Lot NumberW75839-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer12/29/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date01/20/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-