No product was received as the involved device was disposed of at the user facility.Multiple radiographs were provided confirming alleged incident.Review of the reported event identified the patient attempted to climb through a window of her house suggesting that to be the root cause of the device migration.No additional investigation required.Label review: potential adverse events and complications - as with any major surgical procedures, there are risks involved in spinal/orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries." " potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "post-operative warnings - during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." device disposed of at facility.
|