| Catalog Number |
07.01873.006 |
| Medical Device Problem Codes |
Fracture (1260); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the locking cap on an optio-c plate "bent or cracked" upon locking; the doctor chose to leave the plate implanted.There was no delay or impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: product was not returned, however photos were provided which demonstrate the failure.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: per dhr review, the part was likely conforming when it left zimmer biomet control.Device use: this device is used for treatment.
|
| |
|
Search Alerts/Recalls
|