• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CURRENT PLUS VR, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CURRENT PLUS VR, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1211-36Q
Medical Device Problem Code Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code Arrhythmia (1721)
Date of Event 03/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced ventricular tachycardia (vt) and was admitted to the hospital.The patient received external defibrillation to resolve the vt.Device interrogation revealed that the implantable cardioverter defibrillator was in backup vvi.The device was explanted and replaced due to battery exhaustion.The patient was stable post procedure.
 
Additional Manufacturer Narrative
The reported event of backup operation was confirmed.Analysis of device image found the device went into backup mode due to excessive charges resulting in a power on reset (por).After the device was restored from backup mode, functional testing was performed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.A longevity calculation was performed and found the battery depletion was normal based on the device usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT PLUS VR, DF-4 CONNECTOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14136838
Report Number2017865-2022-08185
Device Sequence Number17703809
Product Code NVZ
UDI-Device Identifier05414734503716
UDI-Public05414734503716
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Model NumberCD1211-36Q
Device Catalogue NumberCD1211-36Q
Device Lot NumberS000063585
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/07/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;