| Model Number |
1294-03-030 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Pocket Erosion (2013); Ambulation or Postural Difficulties (2544)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Rupture of the femoral component with femoral bone erosion and fracture of the medial and lateral cortex.Revision surgery necessary.Surgeon refers that the patient had severe hypokinetic syndrome due to progressive inability to walk.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the device associated with this report was returned.Visual examination of the device confirmed the fracture condition.Additionally, debris can still be seen on the device.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this 129403030/2194850 combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Device history review : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this 129403030/2194850 combination.
|
| |
|
Search Alerts/Recalls
|