MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LCS COMPLETE FEM POR R MED; LCS COMPLETE : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LCS COMPLETE FEM POR R MED; LCS COMPLETE : KNEE FEMORAL Back to Search Results
Model Number 1294-03-030
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pocket Erosion (2013); Ambulation or Postural Difficulties (2544)
Date of Event 03/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Rupture of the femoral component with femoral bone erosion and fracture of the medial and lateral cortex.Revision surgery necessary.Surgeon refers that the patient had severe hypokinetic syndrome due to progressive inability to walk.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was returned.Visual examination of the device confirmed the fracture condition.Additionally, debris can still be seen on the device.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this 129403030/2194850 combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Device history review : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this 129403030/2194850 combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMPLETE FEM POR R MED
Common Device Name
LCS COMPLETE : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14137126
Report Number1818910-2022-06879
Device Sequence Number12323666
Product Code NJL
UDI-Device Identifier10603295021681
UDI-Public10603295021681
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/11/2011
Device Model Number1294-03-030
Device Catalogue Number129403030
Device Lot Number2194850
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer05/24/2022
06/27/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/25/2022
06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LCS COMP RP INSERT MED 12.5MM; MBT POR TIBIAL TRAY SZ2
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-