DEPUY SPINE INC. VIPER PRIME PALM HANDLE; SURGICAL INSTRUMENT HANDLE, NON-TORQUE-LIMITING Back to Search Results
Model Number 286750036
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from depuy synthes reports an event in (b)(6) as follows: this is report 1 of 2 for (b)(4).It was reported by the customer in (b)(6) that during a spinal fusion surgery on (b)(6) 2022, it was observed that the handle from the driver was damaged that the handle was not able to be placed on the screwdriver.It was further reported that the stylet depth adjuster end ring was damaged and completely came off while trying to remove the bent stylet from the screwdriver.Another like devices were used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This report is for one (1) viper prime palm handle device.This complaint involves two (2) devices.
 
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The lot number was unknown.Therefore, the expiration date, manufacturing site name and device manufacture date were unknown.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for viper prime palm handle was conducted identifying that lot number mf4341502 was released in one batch.¿ batch1: lot qty of 139 units were released on march 7, 2019 with no discrepancies.Supplier: depuy : metal craft machine & engineering as a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the viper prime palm handle (p/n: 286750036, lot: mf4341502) had multiple scratches around the surface.Also, the handle core component had impaction marks that could be attributable to the unable to assemble condition reported.No other issues were identified.A functional test was performed on the complaint piece with its returned mating device the prime stylet depth adjustor.Both devices cannot be assembled due to the handle core component was slightly deformed.A dimensional inspection was not performed for the viper prime palm handle due to post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper prime palm handle (p/n: 286750036, lot: mf4341502) would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME PALM HANDLE
Common Device Name
SURGICAL INSTRUMENT HANDLE, NON-TORQUE-LIMITING
Manufacturer (Section D)
DEPUY SPINE INC.
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14137321
Report Number1526439-2022-00583
Device Sequence Number19906773
Product Code OLO
UDI-Device Identifier10705034507637
UDI-Public10705034507637
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K170937
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number286750036
Device Catalogue Number286750036
Device Lot NumberMF4341502
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-