| Catalog Number |
0220180518 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was foreign materials inside the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign materials inside the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: debris was inside the package salesforce case number (b)(4) the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be manufacturing.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Search Alerts/Recalls
|