| Model Number |
1804225-28 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion with heavy calcification and moderate tortuosity in the right coronary artery (rca).Pre-dilatation was performed successfully.The lesion was distal to a previously implanted stent.The 2.25 x 28 mm xience skypoint stent delivery system (sds) was advanced, but could not cross the previously implanted stent and the xience skypoint stent dislodged from the balloon.The stent was compressed against the vessel with another xience skypoint stent.Another xience skypoint stent was used to complete the procedure successfully.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported difficulties and subsequent treatments appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|