• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804225-28
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/18/2022
Event Type  Injury  
Event Description
It was reported that the procedure was performed to treat a lesion with heavy calcification and moderate tortuosity in the right coronary artery (rca).Pre-dilatation was performed successfully.The lesion was distal to a previously implanted stent.The 2.25 x 28 mm xience skypoint stent delivery system (sds) was advanced, but could not cross the previously implanted stent and the xience skypoint stent dislodged from the balloon.The stent was compressed against the vessel with another xience skypoint stent.Another xience skypoint stent was used to complete the procedure successfully.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported difficulties and subsequent treatments appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14137349
MDR Text Key289458458
Report Number2024168-2022-04134
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648233029
UDI-Public08717648233029
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/05/2023
Device Model Number1804225-28
Device Catalogue Number1804225-28
Device Lot Number1082641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-