| Model Number |
MS1-5590S |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that the rod has not lengthened at all since implantation.The surgeon's course of action is unknown at this time.No patient adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation as it remains ins-situ.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The rod was not returned for investigation and remains in-situ.During evaluation of the provided x-ray images, it was observed that the rod has not lengthened at all, which confirmed the reported failure mode.The exact root cause of the failure could not be determined due to no device return and no indication of specific failure within the x-ray images.Based on confirmed failure mode and previous similar complaints, it is possible bending forces applied to the rod from patient anatomy/activity may have caused the distraction rod to wedge into the housing tube, which could then cause the rod to become jammed.Based on the complaint description, the surgeon's course of action is unknown at this time.Review of the device history records indicated the unit met all quality specifications and testing prior to product release.If additional information becomes available, a supplemental report will be submitted.H3 other text: x-ray analysis only.
|
| |
|
Event or Problem Description
|
|
No additional event information was provided.
|
| |
|