NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MS1-5590S
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Event or Problem Description
Information was received that the rod has not lengthened at all since implantation.The surgeon's course of action is unknown at this time.No patient adverse event was reported.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation as it remains ins-situ.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
The rod was not returned for investigation and remains in-situ.During evaluation of the provided x-ray images, it was observed that the rod has not lengthened at all, which confirmed the reported failure mode.The exact root cause of the failure could not be determined due to no device return and no indication of specific failure within the x-ray images.Based on confirmed failure mode and previous similar complaints, it is possible bending forces applied to the rod from patient anatomy/activity may have caused the distraction rod to wedge into the housing tube, which could then cause the rod to become jammed.Based on the complaint description, the surgeon's course of action is unknown at this time.Review of the device history records indicated the unit met all quality specifications and testing prior to product release.If additional information becomes available, a supplemental report will be submitted.H3 other text: x-ray analysis only.
 
Event or Problem Description
No additional event information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise #100
aliso viejo, CA 92656
MDR Report Key14137614
Report Number3006179046-2022-00177
Device Sequence Number3986439
Product Code PGN
UDI-Device Identifier00812258026318
UDI-Public812258026318
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMS1-5590S
Device Lot Number0032509AAB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
Patient Weight41 KG
-
-