MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. LINER STANDARD 32 MM I.D. FOR USE WITH 50/52/54 MM O.D. SHELLS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. LINER STANDARD 32 MM I.D. FOR USE WITH 50/52/54 MM O.D. SHELLS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number N/A
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374); Muscle/Tendon Damage (4532)
Date of Event 03/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: ref 00-7713-011-00 lot 63348766 stem, ref 00-6250-65-30 lot 63401075, 63410021 screw x2, ref 00-8775-032-02 lot 2845633 head ¿ biolox, ref 00-6202-54-22 lot 63365704 shell, ref 00-7848-023-00 lot 63330691 kinectiv neck.The device will not be returned for analysis, as the device was discarded; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported the patient underwent right initial total hip arthroplasty.Subsequently was revised approximately 5.5 years later due to external trauma causing the hip to dislocate.The head, neck, and liner were exchanged without complications.The stem and cup remained implanted.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the patient was then revised due to trauma causing the hip to dislocate.Noted hole in capsular tissue.Liner removed and noted to have mild wear in the area where the head was dislocating but otherwise in good condition.The head, neck, and liner were exchanged without complications.Review of the device history records identified no (related) deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No additional information is available regarding the incident.
 
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).The following sections have been updated: a3, b4, b5, d2, d4, g1, g3, g6, h1, h2, h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER STANDARD 32 MM I.D. FOR USE WITH 50/52/54 MM O.D. SHELLS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14137627
Report Number0002648920-2022-00096
Device Sequence Number4293304
Product Code LPH
UDI-Device Identifier00889024123755
UDI-Public(01)00889024123755(17)210228(10)63292496
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K990135
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberN/A
Device Catalogue Number00630505032
Device Lot Number63292496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer05/12/2022
06/18/2024
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/13/2022
06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
Patient Weight92 KG
-
-