Model Number P10.7-80D305
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and is pending evaluation.The root cause cannot be confirmed at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that during explantation the nail disassociated and the crown came off.All fragments of the nail were explanted and the patient outcome was successful.
 
Additional Manufacturer Narrative
Device evaluation: the returned device was inspected and the anti-rotation lug was found to be disengaged from the housing tube which confirmed the reported failure mode.The barbs of the anti-rotation lug had noted damage likely from the disengagement which occurred during the removal of the device.The work order was reviewed, and confirmed the device passed all inspections per the acceptance tests.The manufacturing x-ray image showed the anti-rotation lug (crown) was seated properly in the housing body.Based on the provided information, the damage to the nail most likely occurred during the removal procedure as the patient was able to complete lengthening treatments.The nail experienced excessive torsional force during the removal procedure.Device records review: a review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14137683
Report Number3006179046-2022-00180
Device Sequence Number19869117
Product Code HSB
UDI-Device Identifier00812258020668
UDI-Public812258020668
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K191336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberP10.7-80D305
Device Lot Number0102910AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age18 YR
Patient SexMale
Patient Weight68 KG
-
-