• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. JELCO PROTECTIV SAFETY IV CATHETERS; CATHETER, INTRAVASCULAR, THERAPEUTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. JELCO PROTECTIV SAFETY IV CATHETERS; CATHETER, INTRAVASCULAR, THERAPEUTIC Back to Search Results
Model Number 305606
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of catheter broke inside of a patient.We were able to remove the catheter, but not without causing extreme pain to the patient.
 
Additional Manufacturer Narrative
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could not be duplicated.It was found to be use error.The device passed all functional tests.Dhr review was done, no issues related to the original complaint were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO PROTECTIV SAFETY IV CATHETERS
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14137717
Report Number3012307300-2022-06552
Device Sequence Number6231365
Product Code FOZ
UDI-Device Identifier10351688071101
UDI-Public10351688071101
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K990236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/23/2024
Device Model Number305606
Device Catalogue Number305606
Device Lot Number4205683
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
Patient Weight142 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-