| Model Number |
100/870/080CZ |
| Medical Device Problem Code |
Blocked Connection (2888)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the suction line is clogged.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: device evaluation: one used decontaminated sample 007/910/880 8mm blu suctionaid sub-assy was returned for investigation without its original packaging (see photo of sample).Under visual inspection we noticed that suction tubing has tip formed end in suction connector which is causing reported occlusion.Each suction line assembly shall be inspected as per pwi-10001168 rev.105 operator quality check no.4.2: "flow test each sub-assembly, place the tapered end of the connector on to the test nozzle, and visually confirm that the top of the flow indicator is above the minimum flow requirement of 4 ltr/min @ 70 mbar pressure." testing was replicated on returned sample by manufacturing quality engineer (b)(6) (flow meter g:01874-cz).It was found that flow is occluded and do not meet minimum air flow.The customer stated problem was duplicated.The cause of the reported problem was traced to the manufacturing process.Any other similar customer complaint has not been received against smiths medical czech republic suction line production therefore this incident is considered to be isolated failure only.All operators who are working on affected workplace were informed by manufacturing quality engineer (b)(6).Further more quality alert which is describing what was happened was placed on affected workplace.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.
|
| |
|
Search Alerts/Recalls
|