| Model Number |
N/A |
| Medical Device Problem Codes |
Calibration Problem (2890); Output Problem (3005); Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint has been recorded under (b)(4).Once completed a supplemental/final report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to surgery the device had calibration issues.Once in surgery the device was noted to have different readings and pressure was believed to be too high.No harm or delay were reported along with this event.No other adverse events were reported along with this event.
|
| |
|
Event or Problem Description
|
|
No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint has been submitted under (b)(4).Review of the most recent repair record identified no related findings/repairs to the reported event.Device was assessed and met all specifications.Device was returned to the customer.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|