• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PRECISION IE 4K EYEPIECE ARTHROSCOPE, AUTOCLAVABLE, 4MM X 70°, 140MM, SPEED-LOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PRECISION IE 4K EYEPIECE ARTHROSCOPE, AUTOCLAVABLE, 4MM X 70°, 140MM, SPEED-LOCK Back to Search Results
Model Number 0502-444-070
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
The device manufacture date is not known.This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure "distal tip cracked" was confirmed.According to henke: scope has been evaluated as a level 2 repair due to the distal use, dent/scratches and broken internal optics.Probably root cause for this failure is: the complaint has been confirmed and has been determined to be from customer use and handling.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION IE 4K EYEPIECE ARTHROSCOPE, AUTOCLAVABLE, 4MM X 70°, 140MM, SPEED-LOCK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14138324
Report Number0002936485-2022-00192
Device Sequence Number4728253
Product Code HRX
UDI-Device Identifier07613327405910
UDI-Public07613327405910
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K183470
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502-444-070
Device Catalogue Number0502444070
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-