Visual inspection of returned implants confirms tulip disengagement.The screw shank bulb has multiple witness marks, trending to one side of the tulip head, in close proximity to each other caused by the associated rod.Witness marks indicate that the blocker was tightened multiple times or, more likely, caused by the "helicoptering" of the screw.The head of the shank is stripped.All corners of the hex that engages with the oasys screwdriver(s) are stripped and rounded.The screw shank hex corners deformation confirms what was reported, that the screw stripped.Once screw heads are stripped the polyaxial screwdriver or locking screwdriver would not be able to engage with the screw head to further advance or remove the screw.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Note that "helicoptering" is an off-label technique that involves using a spare piece of rod and a spare blocker to remove the affected screw when unable to use the indicated screwdriver for screw removal.The rod is used to increase applied torque to the screw and the blocker is used to hold the rod in place during removal.This technique is not indicated in the stg.Complaint history was reviewed and no similar complaints for the reported lot number were identified.Initial report states that during a posterior spinal fusion two oasys 4.5 mm x 36 mm screws stripped and the tulips came apart from the screw shank.It was reported via follow up communication that the surgeon had to helicopter both screws out.The first screw tulip disengaged on the back table while the scrub tech was attempting to remove the set screw from the tulip.The second screw tulip disengaged in the patient, likely during helicoptering.It was unable to be determined which of the two returned screws disengaged in the patient or on the back table.The screw holes were prepped using a 3.5 mm tap.It was reported that the screws were not inserted at a difficult angle and excessive force was not applied.The patient had dense bone quality and did not experience any harm from the event, which resulted in a 30 min surgical delay.Device stg reviewed: the non-biased angle polyaxial screws are available in diameters of 3.5mm, 4.0mm and 4.5mm.The 3.5mm and 4.0mm screws are 10 to 20mm in length in 2mm increments while the 4.5mm screws are 20 to 52mm in length in 4mm increments.The non-biased angle polyaxial screws offer up to 30° of angulation in each direction or 60° conical.It is recommended that a tap be used to finalize the preparation of the screw pathway.Taps are available in two lengths, 18mm and 54mm, and three diameters, 3.5mm, 4.0mm, and 4.5mm.There are two types of taps - cancellous and cortical - to match the thread pitch of the screws being used.Note: using incorrect tap or drill combination can result in reduced bone purchase or possibility of instrument breakage.Note: the 4.0mm drill followed by the 4.5mm tap should be used to prepare the pathway for the 4.5mm non-biased angle screws.The most likely cause of the reported event was determined to have two main contributing factors: 1) under-tapping of screw holes coupled with patients dense bone quality.It was reported that the screw holes were prepped by using a 3.5 mm tap.Stg states that "the 4.0mm drill followed by the 4.5mm tap should be used to prepare the pathway for the 4.5mm non-biased angle screws." this most likely caused screw heads to strip during insertion from excessive torque, resulting in the need to helicopter screws out.2) helicoptering screws out is an off-label technique that can induce torques and stresses on the screw that the tulip and shank are not designed or indicated for.It is likely that this technique damaged the tulips and/or shanks which caused the tulips to disengaged either in the patient or on the back table while attempting to remove set screw.
|