| Model Number |
CF-Q160S |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Follow up communication with the customer stated that the endoscope was used in the operating room without endoscopy nursing staff.It came back from the or (operating room) clogged and according to the reporter , it could not be brushed free during pre-cleaning and was sent in for repair that day.The reporter did not report any patient harm or infection , stated that the endoscopes are checked quarterly by the hygiene department and there were no abnormalities.The subject device was received and evaluated.Device inspection and evaluation observed foreign material exiting from the channel ( looks like a white residue deposits on the scope brush) when removed from the channel.The a/w channel was noted to be clogged.Suction channel does not work (clogged) due to blockage with foreign material.Additionally, the following findings found on the device were noted below: connector ( el-connector unit ) found corroded.Control unit painting peeled off.Suction cylinder control unit nut painting peeled off.Scope control unit body , grip , connecting tube pocket pouch has scratch.A-rubber glue (bending section) separated.Distal end cap c-cover has sharp edge (snap).Ccd unit cover lens , light guide rod lens has scratch.Based on device evaluation findings, the customer reported issue of suction does not work was confirmed as during device inspection the suction channel (a/w) unit was found not working.It likely the suction problem was due to clogging found on the channel caused by foreign material.Supplemental report(s) will be submitted should any relevant new information is available.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
| |
|
Event or Problem Description
|
|
Customer sent a repair request reported with a complaint description of suction function does not work.The issue found during reprocessing.There was no patient harm or injury reported due to the event.No user injury reported.Device evaluation found foreign material exiting in the device channel.The a/w (air /water) channel was found clogged due to foreign material blocking the channel.This report is being submitted for foreign material found in the device channel , a/w channel clogged with foreign material found during device evaluation (reportable malfunction foreign material exiting from the channel, insufficient reprocessing).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to g3 of the initial medwatch.The aware date should be (b)(6) 2022.This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.It has been over 11 years, since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been, due to the following: because of incorrect reprocessing after the previous procedure, foreign material remained in the channel.And the subject device was used in such status for inspection, prior to use by the user was insufficient.The channel was clogged, because the user was suctioned the hard material at the procedure.The user can detect the suggested event properly by handling the device in accordance with the following instructions for use (ifu): "inspection of the instrument channel, inspection of the suction function".The user can reduce/prevent occurrence of the suggested event by handling the device in accordance with the following ifu: "if the endoscope is not immediately precleaned, residual organic debris will begin to solidify and it may be difficult to effectively reprocess the endoscope.Brushing the channels: avoid aspirating solid matter or thick fluids, channel or valve clogging can occur.If the suction valve clogs and suction cannot be stopped, disconnect the suction tube from the suction connector on the endoscope connector.Turn the suction pump off, detach the suction valve and remove solid matter or thick fluids".Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|