MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035421030
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual inspection: the proximal end of the coil delivery wire was found to be kinked/bent.The main coil was found to be stretched and broken/fractured.Functional inspection: coil in catheter friction functional test ¿ unable to perform due to condition of the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The device was analyzed.The main coil was found to be stretched and broken/fractured.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Therefore, an assignable cause of procedural factors will be assigned to the as reported code 'coil in catheter friction', and to the as analyzed codes 'main coil stretched', and 'main coil broken/fractured during use'.An assignable cause of 'handling damage' will be assigned to as analyzed code 'coil delivery wire kinked/bent' as these defects appear to be associated with handling of the product or portion of the product during the procedure.
 
Event or Problem Description
On analysis of the returned subject coil, the proximal end of the subject coil delivery wire was found to be kinked/bent and the main coil was found to be stretched and broken/fractured.There were no clinical consequences to the patient reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14139768
Report Number3008881809-2022-00183
Device Sequence Number11758759
Product Code HCG
UDI-Device Identifier04546540697974
UDI-Public04546540697974
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2024
Device Model NumberM0035421030
Device Catalogue NumberM0035421030
Device Lot Number22882883
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Initial Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR MICROCATHETER (STRYKER)
Patient SexUnknown
-
-