MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRACLIP DUAL TRIGGER; BREAST TISSUE MARKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRACLIP DUAL TRIGGER; BREAST TISSUE MARKER Back to Search Results
Model Number 864017D
Medical Device Problem Codes Positioning Failure (1158); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a breast biopsy procedure, the clip was allegedly detached from the support.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 02/2024).
 
Event or Problem Description
It was reported that during a breast biopsy procedure, the clip was allegedly detached from the support.It was further reported that clip had positioning failure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraclip dual trigger breast tissue marker was returned for evaluation.During visual evaluation, the device appeared to have blood residue on the needle and the device was noted to be fully deployed.The pva polymer appeared returned but the wire-form was not returned.There were no visual anomalies noted on the device.Further, functional testing was not performed due to the condition of returned device.Therefore, the investigation is confirmed for the both reported detachment and positioning failure as the device was noted to be fully deployed & the pva polymer returned as the interwoven fiber detaching from the marker.A definitive root cause for the alleged detachment and positioning failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 02/2024), h6 (device).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRACLIP DUAL TRIGGER
Common Device Name
BREAST TISSUE MARKER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14139795
Report Number2020394-2022-00302
Device Sequence Number3986509
Product Code NEU
UDI-Device Identifier00801741032257
UDI-Public(01)00801741032257
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K090547
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number864017D
Device Catalogue Number864017D
Device Lot NumberHUFQ0750
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-