• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number 20226
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Used the first day of the month of the bsc aware date since event date was not provided.
 
Event or Problem Description
It was reported that stent damage occurred.Upon unpacking of an 8.0x30x135cm express ld iliac stent, the proximal stent struts were found to be damaged.The procedure was completed with a different device.No patient complications were reported.
 
Event or Problem Description
It was reported that stent damage occurred.Upon unpacking of an 8.0x30x135cm express ld iliac stent, the proximal stent struts were found to be damaged.The procedure was completed with a different device.No patient complications were reported.
 
Additional Manufacturer Narrative
(b3) date of event: used the first day of the month of the bsc aware date since event date was not provided.Device evaluated by mfr: a express ld was returned for analysis.A visual and tactile examination identified that the balloon was tightly folded and had not been subject to positive pressure.No issues were noted which may have potentially contributed to the complaint incident.An examination of the crimped stent identified that the proximal end of the stent was damaged where the stent struts had been pushed distally.A visual examination found the stent had moved in a distal direction.A visual and tactile examination identified no issues with the shaft of the device.A visual and tactile examination identified no issues with the tip of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14139895
Report Number2134265-2022-04536
Device Sequence Number4142453
Product Code NIO
UDI-Device Identifier08714729579700
UDI-Public08714729579700
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K133110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2024
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0027897844
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-